Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts in the silicone overmold on the top cover.In addition, the electrode had multiple cuts prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed cut electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|