On march 15, richard wolf mic received medwatch report mw5099373 with the following event description: the manufacturer's instructions for use (ifu) is inadequate.The device is unable to be thoroughly cleaned.Residual bioburden is found in the shaft of the probe and inside the cone.The manufacturer/model numbers are as follows: richard wolf: intrauterine probe - 8378.00, small cone - 8378.90, large cone - 8378.91.Boroscope inspection revealed what appeared to be biofilm colonies.Please also see mdr 1418479-2021-00014 and 1418479-2021-00013.
|
|
Richard wolf medical instruments (rwmic) is submitting this mdr on behalf of the device manufacturer, rw gmbh.Follow-up report #1 is to provide fda with new and corrected informations about this mdr.The following fields have: new information: b4, b5, d8, e1, e2, e3, g6, h2, h7, and h10.Corrected information: d4, g3, h3, h6-investigation findings.Results of the device product history evaluation: the device was not returned instead, a product history evaluation was performed.According to the investigation report product history evaluation , form 5253 (00): (04/19/2021) rwgmbh has not received similar complaints about the product 8378.00.In the last 10 years the only one complaint was from rwmic (19-00178) but with another problem description: "tip inside the patient".There is no further information in the complaint records.Based on the available results, there are no signs of systematic problems.No further action is required.Rwgmbh has revised the device instructions for use (ifu) as a result of this complaint at rwmic.Ifu number: ifu ga-e074 en us.Old revision no.: index: 2012-01 v1.0 / eco 2011-0351.New revision no.: v2.0 / 2022-04 / pk22-0448.Ifu released date: 12-july-2022.In general, the user is advised in the associated instructions for use ga-e074 en us under chapter 9 (reprocessing and maintenance) that a visual and functional check must be carried out before and after each use, as well as during reprocessing.Notice: (important).Observe the following when cleaning the products: - products with long channels: pull through the cleaning brush.Insert the cleaning brush into the channel with the brushless side first and pull out the brush on the other side of the channel.- other products/parts: use a brush pass move the cleaning brush forwards and backwards once.Specifically, section 9.1, describes fully the reprocessing sequences for manual and machine cleaning (in w/d) processes.Richard wolf medical instruments corporation (rwmic) considers this mdr closed.Rwmic will submit a follow up report if new information becomes available.
|