| Model Number |
354807 |
| Medical Device Problem Codes |
Failure to Capture (1081); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Lead remains implanted.Representative reported oversensing on the lv channel.Lv iegm is not visible on transmitted recordings, however the second lvs event appears to line up with the t-wave on the far-field channel.Loss of capture was also reported.Lead to be evaluated further.No adverse patient events reported.Should additional information become available, it will be added to this file.
|
| |
|
Additional Manufacturer Narrative
|
|
As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.
|
| |
|