| Model Number |
72202618 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during rotator cuff tendon repair, the anchor came out from bone hole when tying the sutures.The procedure was completed using loose sutures with needles.It is unknown if a delay occurred.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during rotator cuff tendon repair, the anchor came out from bone hole when tying the sutures, the anchor was removed.The procedure was completed with needle sutures, fixing the tendon to the bone.No additional bone holes were required.20 minutes of delay occurred.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|