SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568S |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: case-(b)(4).
|
|
Event Description
|
It was reported that, during setup for a shoulder procedure, the light source powered off unexpectedly.It is unknown if there was a backup device available and there was no delay reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event, include poor ventilation due to external obstructions to vents or excessive dust buildup on rear fans or fan guards.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
|
Manufacturer Narrative
|
H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.Complaint of power loss could not be reproduced.Product passed functional testing and 2 hour burn-in with no faults or errors.All functions perform as expected.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.It was determined the device did not contribute to the reported event.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event, include overheating of the power supply caused by poor ventilation due to external obstructions to vents.No containment or corrective actions are recommended at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
|