Model Number ENBF3620C166E
Medical Device Problem Codes Leak/Splash (1354); Material Deformation (2976)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Other relevant device(s) are: product id: enlw1620c124e, serial/lot #: (b)(4), ubd: 15-may-2013, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant stent graft system was implanted in an unknown endovascular procedure.It was reported by the patient through technical services, that approximately 7 year post index procedure, a repair was performed to the previous implanted stents to straighten them back out.As per patient, it was found out that instead of an upside-down "y" shape, the stents gotten into a "c" shape.The patient does not think the stents grafts were explanted and does not know if new stent grafts were implanted, but notes it is thought they were "rewrapped".No cause of the event was reported.No additional clinical sequalae were reported.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
It was reported that bilateral limb extensions (non medtronic) were implanted during the intervention in 2018 to treat a type 1b bilateral endoleak.It was reported per the physician that the cause of the event was type 1b endoleak and enlarging of the aaa sac.It was reported that there was no device malfunction; rather distal artery dilitation resulting in loss of distal seal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key11676763
Report Number9612164-2021-01486
Device Sequence Number3415548
Product Code MIH
UDI-Device Identifier00613994762351
UDI-Public00613994762351
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/05/2013
Device Model NumberENBF3620C166E
Device Catalogue NumberENBF3620C166E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/19/2021
Supplement Date Received by Manufacturer08/05/2021
08/13/2021
Initial Report FDA Received Date04/16/2021
Supplement Report FDA Received Date08/11/2021
08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-