| Model Number |
LC358L |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been made available for evaluation.Should the device become available, or further information be provided, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Provider alleges consumer got stuck in the recline position and had to call the paramedics to get him out of the chair.
|
| |
|
Additional Manufacturer Narrative
|
|
Only the actuator was returned for evaluation.The actuator failed to reach full extension during load testing and would not retract.This functional failure was identified as a worm gear failure.It is unknown what caused this gear to fail.
|
| |
|
Event or Problem Description
|
|
Provider alleges consumer got stuck in the recline position and had to call the paramedics to get him out of the chair.
|
| |
|