MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97714
Device Problems Electromagnetic Interference (1194); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Pain (1994); Burning Sensation (2146); Electric Shock (2554)
Event Date 04/13/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 97715, serial#: (b)(4), implanted: (b)(6) 2018, product type: implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for spinal pain.It was reported that the patient reported to clinic with issues after an mri.Patient stated she felt burning/stabbing shortly after mri started, and it had to be aborted.The patient reported painful zap/burn mid back and upwards to neck.The tech reported running 1.5 and sar at < 2 watts/kg.The patient was confirmed to be implanted with mri safe systems.The tech reported using mri guides for patient's specific device.The tech also stated that the patient has had similar mris in the past with them using the same criteria.Troubleshooting/diagnostics were performed.Impedances were tested and performing within range.The devices were determined to be fine and in service.The issue was resolved at the time of the report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11678553
MDR Text Key245816650
Report Number3004209178-2021-06149
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2017
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2021
Date Device Manufactured03/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
-
-