| Model Number |
72202087 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip damage, fiber damage, and a cracked and scratched distal lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution. a complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that, the "hd arthroscope (4mm x 30°)" showed a halo around the picture.No case reported; therefore, there was no patient involvement.Results of investigation have concluded that this unit had a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3,h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip damage, fiber damage, and a cracked and scratched distal lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.
|
| |
|
Search Alerts/Recalls
|