ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550300-28
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Enzyme Elevation (1838); Prolapse (2475)
Date of Event 09/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.It should be noted that the reported patient effect of prolapse is listed in the xience sierra, everolimus eluting coronary stent system, instructions for use, as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined; however, the reported treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
Patient id: (b)(6).It was reported that on (b)(6) 2020, a percutaneous intervention was performed on the distal right coronary artery (rca), 95% stenosed lesion.Pre-dilatation was performed and a 3.0x28mm xience sierra stent was implanted in the distal rca, without a device issue reported.Following, intra-stent plaque protrusion in the distal stent segment was observed, treated with additional dilatation.Timi flow iii and 0% residual stenosis was observed.Elevated cardiac enzymes were observed post-procedure.There was no myocardial infarction, no treatment, and no serious association with the elevated enzymes.The patient was discharged home the same day the elevation was observed.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11679143
Report Number2024168-2021-03208
Device Sequence Number14233393
Product Code NIQ
UDI-Device Identifier08717648227288
UDI-Public08717648227288
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/02/2021
Device Model Number1550300-28
Device Catalogue Number1550300-28
Device Lot Number0052741
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/29/2021
Initial Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight99
-
-