ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000320
Device Problem Gas/Air Leak (2946)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 03/30/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
There was no patient involvement.It was reported that helium leakage detected during preventative maintenance (pm).
 
Event Description
There was no patient involvement.It was reported that helium leakage detected during preventative maintenance (pm).
 
Manufacturer Narrative
Qn# : (b)(4).No iabp part or recorder strip was returned to teleflex chelmsford for investigation.The reported complaint of helium leakage was confirmed by the field service agent.According to the eqr, the pcs assembly was replaced.If the product is returned at a later date, a full investigation of the sample will be completed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11679715
MDR Text Key248027174
Report Number3010532612-2021-00113
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902051715
UDI-Public30801902051715
Combination Product (y/n)N
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN000320
Device Catalogue NumberIAP-0500I
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-