| Model Number |
IPN048895 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A verification of the reported failure mode was conducted, and 400 samples were taken from the current production (material number 00003-14 lot # 3027607) at the manufacturing facility.The samples were visually inspected, and issue reported "leak - device leak - cuff" was not observed in the current manufacturing process.A device history record review could not be conducted since the lot number of the device was not provided.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported "ett connector coming of the tube and there is leakage around the cuff".Device replaced with no patient injury or harm reported.Patient condition reported as "fine".
|
| |
|