• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL PERICARDIAL BIOPROSTHESIS; REPLACEMENT, HEART-VALVE Back to Search Results
Model Number 7000TFX
Medical Device Problem Code Gradient Increase (1270)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.The root cause of this event cannot be conclusively determined with the available information.However, the stenosis, in this case, was most likely impacted by the progression of the patient's underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.The subject device is not available for evaluation, as it remains implanted in the patient.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, an appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 25mm 7000tfx mitral valve underwent a valve-in-valve after an implant duration of 10 years, (3) months due to stenosis.The tmvr was performed with a 26mm 9750tfx.
 
Additional Manufacturer Narrative
H11: corrective data: based on the additional information obtained, this event is no longer considered reportable and this correction is being submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key11680342
Report Number2015691-2021-02510
Device Sequence Number11616808
Product Code DYE
UDI-Device Identifier00690103167849
UDI-Public00690103167849
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
NI
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2013
Device Model Number7000TFX
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/22/2021
Supplement Date Received by Manufacturer05/11/2021
Initial Report FDA Received Date04/16/2021
Supplement Report FDA Received Date05/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age78 YR
-
-