Catalog Number 305618 |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/18/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the syringe 30ml ll tip conv pak experienced packaging tears while opening leaving paper shards.The following information was provided by the initial reporter: material no: 305618, batch no: 0307723.Upon opening the sterile package, a big black piece of fuzz/cardboard fell off the top paper liner and fell into the container with the syringes.This is the 5th package of 30 ml syinges that we have found foreign debris.
|
|
Manufacturer Narrative
|
H6: investigation summary: a device history record review was completed by our quality engineer team for provided lot number 0307723.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.
|
|
Event Description
|
It was reported that the syringe 30ml ll tip conv pak experienced packaging tears while opening leaving paper shards.The following information was provided by the initial reporter: material no: 305618, batch no: 0307723.Upon opening the sterile package, a big black piece of fuzz/cardboard fell off the top paper liner and fell into the container with the syringes.This is the 5th package of 30 ml syinges that we have found foreign debris.
|
|
Search Alerts/Recalls
|