• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD LEUCOCOUNT; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD LEUCOCOUNT; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number 341002
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that prior to use with bd leucocount¿ there was a mix of product.The customer received boxes of products 341001 and 341002, the box of 341002 contained vials for rbc control (same as 341001).The following information was provided by the initial reporter, translated from (b)(6) to english: bd leucocount tm plt control ref 341002 lot lp0321 exp 2021-05-05.Data on tubes inner package: vial 1: bd leucocount tm rbc control high rbc lot lr0321h exp 2021-05-05.Vial 2: bd leucocount tm rbc control low rbc lot lr0321l exp 2021-05-05.Observations and conclusions - looks like a mistake in the packaging of the product.The customer received boxes of products 341001 and 341002 as described.In contrast, box 341002 contained vials for rbc control (same as 341001).
 
Event or Problem Description
It was reported that prior to use with bd leucocount¿ there was a mix of product.The customer received boxes of products 341001 and 341002, the box of 341002 contained vials for rbc control (same as 341001).The following information was provided by the initial reporter, translated from polish to english: bd leucocount tm plt control ref 341002 lot lp0321 exp 2021-05-05.Data on tubes inner package: vial 1: bd leucocount tm rbc control high rbc lot lr0321h exp 2021-05-05.Vial 2: bd leucocount tm rbc control low rbc lot lr0321l exp 2021-05-05.Observations and conclusions - looks like a mistake in the packaging of the product.The customer received boxes of products 341001 and 341002 as described.In contrast, box 341002 contained vials for rbc control (same as 341001).
 
Additional Manufacturer Narrative
H6: investigation summary the scope of issue is limited to part 341002, lot lp0321.Manufacturing defect trend: this product is manufactured by a third party.Manufacturing non-conformance data are not routinely available.Complaint trend: there are no complaints related to the reported complaint for the date range between 26mar2020 and 26mar2021.Batch history record (bhr) review: documentation for part 341002, lot lp032 was reviewed.No discrepancies or non-conformances were observed.The materials met all the manufacturing specifications prior to release.Retain sample evaluation / testing: the retain samples for part 341002, lot lp0321 were visually inspected and found unremarkable.The report was not of a product performance issue.Returned sample evaluation: the sample was not requested to be returned because photos were provided.The photos were evaluated and found to support the complaint described.Investigation result / analysis: the investigation was performed by the supplier (#2021-691-1, attached) and the cause assigned to human negligence.Based on the review of complaint trend, defect trend, bhr, root cause and risk analysis, the reported complaint was confirmed.Risk analysis: 10000525149 pfmea, revision 01 was reviewed.Hazard:_missing kit component_____ cause:_operator skipped component when packaging______; harmful effects:_cannot run assay; delay in testing__; severity: _3__; probability: _2__; risk index: _18__ ; new hazard: _no_; mitigation(s) sufficient _yes__.Root cause analysis: based on the investigation result, root cause was determined to be human negligence.Investigation conclusion: based on the investigation result, complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LEUCOCOUNT
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
MDR Report Key11681745
Report Number2916837-2021-00181
Device Sequence Number3495078
Product Code GKZ
UDI-Device Identifier00382903410026
UDI-Public00382903410026
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
BK000036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/05/2021
Device Model Number341002
Device Catalogue Number341002
Device Lot NumberLP0321
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer10/19/2021
Initial Report FDA Received Date04/16/2021
Supplement Report FDA Received Date11/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-