PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Model Number 4006U0203
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pac400) and non-penumbra microcatheters.During the procedure, it was reported that the pac400 would not deploy.Therefore, it was removed.It is unknown how the procedure was completed or if there was any adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned pac400 revealed that the pusher assembly mid-joint was retracted through the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.Forceful advancement against this resistance likely contributed to the pusher assembly kinks near the mid-joint.Further evaluation revealed dried blood within the introducer sheath distal tip.This damage was likely incidental to the reported complaint.During functional testing, the pac400 was unable to be advanced at all from its returned position within the introducer sheath.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400) and non-penumbra microcatheters.During the procedure, it was reported that the pc400 would not deploy.Therefore, it was removed.It is unknown how the procedure was completed or if there was any adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following sections were incorrectly reported on the initial mfr report and are being corrected on this follow-up #02 mfr report: 1.Section b.Box 5.Describe event or problem.2.Section d.Box 1.Brand name h3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11681897
Report Number3005168196-2021-00772
Device Sequence Number3423713
Product Code HCG
UDI-Device Identifier00814548011363
UDI-Public00814548011363
Combination Product (Y/N)Y
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/13/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4006U0203
Device Catalogue Number4006U0203
Device Lot NumberF89903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer05/25/2021
01/14/2005
Initial Report FDA Received Date04/16/2021
Supplement Report FDA Received Date06/18/2021
08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1