| Model Number |
72290007 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder surgery the wired foot pedal was not working.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Visual inspection observed the units paint is chipped.Functional evaluation revealed no issues.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
|
| |
|
Search Alerts/Recalls
|