Model Number 71337546
Medical Device Problem Code Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during thr surgery, the r3 0 degree xlpe acetabular liner 28 mm inner diameter x outer diameter could not be fixed in the cup.After multiple attempts, the device could not be used and a backup device from smith and nephew was used to complete the procedure.The procedure finished with a delay of less than 30 minutes and no injury to the patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.The contribution of the device to the reported event could not be corroborated.A visual inspection of the returned device did not confirm the stated failure mode.No attributes can be visually found to contribute the failure mode.A dimensional evaluation of the returned device could not confirm the stated failure mode.None of the features that could attribute to the failure mode were found to be out of tolerance.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Some potential causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 28MM X 46MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks road
memphis, TN 38116
5123913905
MDR Report Key11683762
Report Number1020279-2021-03173
Device Sequence Number14225939
Product Code MBL
UDI-Device Identifier03596010598455
UDI-Public03596010598455
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71337546
Device Catalogue Number71337546
Device Lot Number19FM02262
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2021
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer07/21/2021
12/29/2021
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date07/22/2021
01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1