SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 28MM X 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
Model Number 71337546 |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/23/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during thr surgery, the r3 0 degree xlpe acetabular liner 28 mm inner diameter x outer diameter could not be fixed in the cup.After multiple attempts, the device could not be used and a backup device from smith and nephew was used to complete the procedure.The procedure finished with a delay of less than 30 minutes and no injury to the patient.
|
|
Manufacturer Narrative
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.The contribution of the device to the reported event could not be corroborated.A visual inspection of the returned device did not confirm the stated failure mode.No attributes can be visually found to contribute the failure mode.A dimensional evaluation of the returned device could not confirm the stated failure mode.None of the features that could attribute to the failure mode were found to be out of tolerance.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Some potential causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|