| Model Number |
VENUM14120 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, image was provided for review.The investigation of the reported event is currently underway.(expiry date: 10/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during stent deployment on bilateral iliac via bilateral popliteal, the stent proximal end allegedly failed to expand.It was further reported that stent got deployed after twisting the deployment system.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the physical sample was not available.X-ray image provided demonstrating an hour-glass like deformation of the stent inside the vessel in the area of the stent brake confirmed that the stent adhered to the inner catheter stent brake during deployment which leads to a confirmed investigation result for expansion issue.A definitive root cause could not be determined based upon the available information.Labeling review: a review of the relevant instructions for use for this product was conducted.The instructions for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 10/2022), g3 h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during stent deployment on bilateral iliac via bilateral popliteal, the stent proximal end allegedly failed to expand.It was further reported that stent got deployed after twisting the deployment system.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|