• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ASCOPE ACLAVE HD 4MM X 30 DEG; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ASCOPE ACLAVE HD 4MM X 30 DEG; ARTHROSCOPE Back to Search Results
Model Number 72202087
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the scope's external lens was not functionally, the external lens was fractured.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).The reported device was returned to the designated complaint unit for independent evaluation.A visual inspection of returned device found outer tube dented and bent, distal tip and fiber damage, loose inner assembly, distal cover glass and negative lens cracked and scratched, rust and discoloration, and missing paint.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review of the part found similar events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of capas, ncs, pra/hhes, and field actions found no prior escalation actions applicable to the complaint.There was a relationship found between the device and the reported event.The complaint was confirmed, and the root cause was associated with an unintended use of the device.Factors that could have contributed to the reported event include inadequate maintenance of the device and an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCOPE ACLAVE HD 4MM X 30 DEG
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11684587
Report Number3003604053-2021-00161
Device Sequence Number3095582
Product Code HRX
UDI-Device Identifier03596010616869
UDI-Public03596010616869
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K962075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202087
Device Catalogue Number72202087
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-