Model Number 399.091
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the spring blade inside the clamp broke.There were no patient consequences.This report involves one (1) bone holding forceps-soft ratchet f/plates to 9mm wide.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.This complaint is confirmed as the device is broken.While a definitive root cause could not be identified, it is possible that an unintended force on the plate could have contributed to the complaint condition.No design issues were observed during the document/specification review.During this investigation no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot: part: 399.091, lot: 5800662, manufacturing site: salzburg, release to warehouse date: 20 june 2008.Due to the age of more than 10 years of the complained device a wear or use related root cause is the most likely reason of the complained malfunction.Per franchise complaint product investigation procedure is for complaints for which a non-manufacturing related probable cause has been identified no manufacturing record evaluation is required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE HOLDING FORCEPS-SOFT RATCHET F/PLATES TO 9MM WIDE
Common Device Name
FORCEPS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11684760
Report Number2939274-2021-01914
Device Sequence Number19586860
Product Code HTD
UDI-Device Identifier10886982202314
UDI-Public(01)10886982202314
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number399.091
Device Catalogue Number399.091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2021
Initial Date Received by Manufacturer 03/23/2021
Supplement Date Received by Manufacturer05/12/2021
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date05/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-