MAUDE Adverse Event Report: RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RADIAL ARTERY PRESSURE MONITORING SET; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Foreign Body In Patient (2687)
Date of Event 03/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter: customer (person): street: - (b)(6).Reporter occupation: administrative supervisor.Report source - other: mw (b)(4) pma/510(k) #: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that the catheter of a radial artery pressure monitoring set broke, leading to an intervention to remove a portion of the device left behind in the patient.A (b)(6) year old male patient underwent a procedure in which the radial artery pressure monitoring set was used.The catheter snapped while dressing was being removed on the patient's arterial line.Staff were unable remove the device due to bleeding from the site.Pressure was held to the site and a stat us was ordered to visualize the catheter.The or was then booked for exploratory surgery in order to remove the catheter.The right hand remained warm with good capillary refill and pulse.The surgeon was able to retrieve the catheter at bedside.Additional information regarding event details and patient outcome have been requested, but is currently unavailable.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
In additional information received on 21apr2021, it was reported that the patient is "doing ok" after beside removal by the vascular surgeon.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: d9 - returned to manufacturer on, d10 - concomitant products: cook device returned with competitor's extension tubing attached.Investigation ¿ evaluation.It was reported by (b)(6) , vice president of legal affairs for (b)(6) hospital, usa via a medwatch form 3500a, that an arterial catheter separated.The patient was reported to be a 39-year-old male at the time of the event.On (b)(6) 2021, while the dressing was being removed from the patient¿s arterial line, the arterial line separated.The separated portion of the catheter remaining in the patient was unable to be retrieved by the provider who was removing the dressing.Pressure was held to the site after the catheter separation.An ultrasound was ordered immediately, and the catheter was visualized.The operating room was scheduled for exploratory surgery to remove the catheter.The patient¿s right hand remained warm with good capillary refill and pulse.A vascular surgeon was able to retrieve the catheter at the bedside.The patient was reported to be doing ¿okay¿ after the event.Reviews of documentation including the complaint history, device history record (dhr), quality control, as well as a visual inspection of the returned device were conducted during the investigation.One used radial artery pressure catheter returned to the manufacturer for evaluation with a competitors extension tubing attached to the catheter.The catheter separated in a short section beyond the hub.The lumen of the catheter shaft attached to the hub appeared to be flattened.Cook has concluded that the device was manufactured to specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot and component lot revealed no non-conformances relevant to the failure mode.A database search found no other complaints have been reported for the complaint lot.Based on the available information, cook has concluded that there is no evidence suggesting nonconforming product exists either in house or in field.Based on the information provided, inspection of the returned device, and the results of the investigation, a definitive root cause for this event could not be established.A clinical assessment of the complaint was unable to rule out care and maintenance of the device, or inadvertent tension placed on the device as a cause of the event.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL ARTERY PRESSURE MONITORING SET
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key11684761
Report Number1820334-2021-01151
Device Sequence Number3085478
Product Code DQO
UDI-Device Identifier00827002019089
UDI-Public(01)00827002019089(17)231213(10)13626746
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2023
Device Model NumberN/A
Device Catalogue NumberC-PMS-300-RA
Device Lot Number13626746
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2021
Initial Date Received by Manufacturer 04/21/2021
Supplement Date Received by Manufacturer04/21/2021
07/23/2021
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date04/22/2021
07/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN MANUFACTURER EXTENSION TUBING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-