Model Number 190713
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2021
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.  .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a battery holder that sparked. The battery was being replaced during preventative maintenance (pm) and the battery holder sparked upon insertion of a new battery. The machine then repeatedly failed the battery self-test. A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction. The biomed replaced the power logic board, which resolved the issue. The machine has approximately 40,000 hours and the power logic board was the original fresenius part on the machine. The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the power logic board. The unit was returned to service at the user facility without issue and without reoccurrence of the event. The power logic board is not available to be returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008T HEMODIALYSIS SYS., WITH CDX
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11685042
MDR Text Key246007797
Report Number2937457-2021-00699
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/30/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number190713
Device Catalogue Number190713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received04/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-