• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number ETBF3616C166E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in the patient for the endovascular treatment of a 51mm diameter aaa.Approximately 7 years later, it was reported the patient returned for follow up where a suspected type ia endoleak was observed.Per the physician the cause of the event could not be determined.No additional clinical sequalae were reported and the patient will be monitored.
 
Additional Manufacturer Narrative
As of just over a month post its diagnosis, no intervention has been performed for the type ia endoleak and the patient continues to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key11685331
Report Number9612164-2021-01523
Device Sequence Number16926027
Product Code MIH
UDI-Device Identifier00613994991041
UDI-Public00613994991041
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2016
Device Model NumberETBF3616C166E
Device Catalogue NumberETBF3616C166E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/25/2021
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-