Model Number MX693R |
Device Problems
Insufficient Flow or Under Infusion (2182); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/06/2021 |
Event Type
malfunction
|
Event Description
|
Information received a interventional imaging|medex high pressure tubing malfunctioned.The report indicated that on two different occasions the tubing ruptured when doing pump pressure injection during a procedure.Rated for 900psi, were using 700-800 psi.No patient injury during procedure was reported.
|
|
Manufacturer Narrative
|
Other text: returned device was received for evaluation.During the evaluation of the device the customer reported condition was confirmed.Conclusion: the defect reported could not be attributable to the manufacturing process, since the 46 samples passed the high pressure test and the reported lot was found within specifications during high pressure test during its manufacture.Problem source is unknown.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
|