MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER
|
Back to Search Results |
|
Model Number PED2-450-14 |
Device Problems
Activation Failure (3270); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received a report that the pipeline would not open in the distal segment.The middle segment of the stent had been positioned in a bend, more than 50% of the device had been deployed, and resheathing was performed more than twice.A continuous flush had been used, and there was no damage observed to the catheter or pushwire.As the resheathing and manipulation of the device did not open the stent, the physician attempted to remove the device.However, the stent became jammed in the proximal end of the microcatheter.Once both devices were removed, the microcatheter was cut open to ensure the stent had not deployed in the guide catheter or elsewhere in the system.A new pipeline and phenom 27 were used to complete the procedure successfully.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Post-procedure angiographic results showed the device was removed with no issues.The patient was undergoing surgery for treatment of a blister-like, ruptured aneurysm of the right para-ophthalmic segment of the internal carotid artery (ica) with a max diameter of 1.5 mm and a 1.2 mm neck diameter.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet therapy (dapt) was administered, and the platelet reactivity units (pru) level was unknown.
|
|
Manufacturer Narrative
|
H3: analysis of the pipeline flew embolization device (lot no.B160754) found that the distal and proximal dps restraints were intact.The dps sleeves were found intact with no signs of damage.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The distal and proximal ends of the pipeline flex braid were found fully opened and moderately frayed.Bends were found at 12.0 cm to 24.0 cm from the proximal end.No defects were found with the tip coil, distal marker, re-sheathing marker, re-sheathing pad or with the proximal bumper.The total and usable lengths of the catheter were measured to be within specifications.The catheter tip and marker were examined; no damages were found.The catheter body appeared to be accordioned at 7.0 cm to 16.5 cm from the distal tip.No flash or voids molded were observed in the hub.The pipeline flex could not be pushed forward or removed.For further examination, the phenom catheter was cut to remove the pipeline flex delivery system.The catheter was flushed with water and found patient.The catheter was then tested by running an in-house 0.0265¿ mandrel through catheter tip and hub.The mandrel successfully passed through the catheter hub with no issues; however, resistance was observed at the damaged locations.No other anomalies were observed.Based on the returned devices, the pipeline flex and phenom 27 catheter were confirmed to have ¿resistance during retrieval¿ and ¿catheter resistance¿ issues as the returned pipeline flex was found stuck inside the phenom catheter.Regarding to the "failure to open", the customer complaint could not be confirmed as the distal and proximal ends of pipeline flex braid were fully opened and frayed.The damage to the braid on the ends of the pipeline flex braid is likely the results of the physician re-sheathing the device more than recommended two times.From the damages seen on the catheter (accordioning), pusher (bending), pipeline flex braid (fraying) and hypotube (stretching); it appears there was high force used.It is likely these damages occurred when the customer attempted to advance/retrieve the pipeline flex through the phenom catheter against the resistance.However, the root cause could not be determined.Possible causes of the resistance during delivery include patient's vessel tortuosity and lack of continuous flush with heparinized saline during procedure.There was no non-conformance to specifications identified that led to the resistance issue.H6: method code updated to b01.Result code updated to c0702 and c070601.Conclusion code updated to d15.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|