• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL PERCEPTA CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL PERCEPTA CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number W1TR01
Device Problems High impedance (1291); Misconnection (1399)
Patient Problem Insufficient Information (4580)
Event Date 03/26/2021
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 6725, adaptor, implanted: (b)(6) 2021.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was "not feeling right" two days post cardiac resynchronization therapy pacemaker (crt-p) change.On interrogation it was noted that there was an alert for high and undefined right ventricular (rv) lead impedance.The pocket was reopened and the rv lead was unscrewed from the device and re-inserted.The impedance values returned to normal and the crt-p and rv lead remain in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA CRT-P MRI SURESCAN
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11685737
MDR Text Key246023723
Report Number9614453-2021-01625
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169891401
UDI-Public00643169891401
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/14/2022
Device Model NumberW1TR01
Device Catalogue NumberW1TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2021
Initial Date FDA Received04/19/2021
Date Device Manufactured11/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
407652 LEAD, 429688 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age76 YR
Patient Weight53
-
-