| Model Number |
NX028Z |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Implant Pain (4561); Swelling/ Edema (4577)
|
| Date of Event |
06/03/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: no product at hand.Therefore a failure description at the device is not possible.Investigation no product at hand.Therefore an investigation at the product is not possible.Batch history review the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Explanation and rationale, conclusion and root cause in the light of the little information received and due the circumstance that we did not received the complained devices it is not possible to determine a root cause for the mentioned failure.Corrective action for this topic (vega loosening), a product safety case (psc) was initiated.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with a vega ps femoral component.As a result of having the product implanted, the patient has experienced knee pain, difficulty walking, swelling, and loosening and instability of the implant.The primary surgery occurred on (b)(6) 2015, and the revision surgery occurred on (b)(6) 2016.A revision was required.All available information has been provided at this time; if additional information becomes available the complaint will be updated accordingly.The adverse event is filed under xc reference (b)(4).Cement used: unidentified.
|
| |
|
Search Alerts/Recalls
|