| Model Number |
500-56150 |
| Medical Device Problem Codes |
Break (1069); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The all msd groups disabled alert was confirmed through event log and device analysis.The evidence of the rf drift on the control unit (cu) likely accelerated wear on the ultra sonic core (usc).The char seen on the usc points to a short and failure of one of the transducer elements in the usc.It appears the shorting in the usc may have contributed to the failure and eventual fracture of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that an ultrasonic core fracture occurred.An ekosonic catheter was selected for use during a deep vein thrombosis procedure.During the procedure an all msd groups disabled alarm occurred.The alarm was unable to be resolved; therefore, the procedure was completed by running the catheter as a standard infusion catheter.Upon recheck the following day, it was discovered that the ultrasonic core wire had broke inside the patient.The catheter was successfully snared out with a non-bsd snare.No patient complications were reported.
|
| |
|