| Model Number |
SW986K |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that there was an issue with sw986k-activ l inf.Plate s1 size m 5°/spikes.According to the complaint description,a postoperative revision occurred due to subsidence.The original implantation was on (b)(6) 2020.Additional information was not provided.The adverse event / malfunction is filed under aag reference (b)(4).Associated components: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: up to now the products were not provided.Therefore, no investigation was possible.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Explanation and rationale/conclusion and root cause: because there are no products and only minor information available, a definitive root cause analysis is not possible.Following causes are possible: - wrong system configuaration selected by the user.- wrong implant size chosen by the user.- end plate formed too stron by the user.- design layout unsuitable.- inadequate patient behaviour.Because the device history record is without deviation, a material defect or manufacturing failure can be excluded.No capa required.
|
| |
|
Search Alerts/Recalls
|