• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Model Number 72203967S
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).The reported device, intended for use in treatment, was returned to the designated complaint unit for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device noted residue on the connector contacts.Functional evaluation revealed that there was noisy image and image loss when the connector is moved.The complaint has been confirmed.The instructions for use (ifu), provided with the device, contain the following warnings and precautionary measures related to proper use of the device, including but not limited to: the product must be cleaned and sterilized before every subsequent use; use only neutral ph cleaners to decontaminate the camera head.Non-neutral ph cleaners may cause residue buildup, which may cause interference in the video output.A review of the device history records showed there were no indications to suggest that the lot did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Event or Problem Description
It was reported that the screen of the camera head was showing purple queue.It is unknown when the issue was discovered, how the procedure was finished, and if a delay occurred.However, no patient injuries were reported.Results of investigation have concluded that this unit had loss of image which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11686458
Report Number1643264-2021-01461
Device Sequence Number2980720
Product Code HRX
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K153606
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2021
Initial Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-