• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURN ART INS BCS STD 7-8 RT11; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURN ART INS BCS STD 7-8 RT11; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74023273
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/22/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, there was a revision surgery for a right journey bcs 7-8 11mm insert on (b)(6) 2021 due to a broken post.
 
Additional Manufacturer Narrative
H6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, reportedly, the patient underwent revision approximately 10-13 years post implantation due to ¿a broken post¿.Per correspondence, the requested medical documentation was not available for this case.Without the requested documentation, the clinical root cause of the reported event could not be fully confirmed nor concluded.The patient impact beyond the reported post breakage and revision could not be determined.No further medical assessment could be rendered at this time.Should clinically relevant documentation and/or product evaluation results become available, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURN ART INS BCS STD 7-8 RT11
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11687106
Report Number1020279-2021-03194
Device Sequence Number14186950
Product Code JWH
UDI-Device Identifier03596010555298
UDI-Public03596010555298
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K042515
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2017
Device Model Number74023273
Device Catalogue Number74023273
Device Lot Number07AM00639
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/22/2021
Supplement Date Received by Manufacturer07/20/2021
Initial Report FDA Received Date04/19/2021
Supplement Report FDA Received Date07/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-