• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752437
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A doctor reported that the probe did not cut and dragged vitreous during a procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of the cutter was not working and it was dragging the vitreous; therefore, the condition of the product could not be verified.The product was processed and released according to the product¿s acceptance criteria.A sample was not received at the manufacturing site and the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
One opened probe was received, with a tip protector, in a tray along with other items.The sample was visually inspected and found to be nonconforming with orange/brownish foreign material on the port face.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for both functional test.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be functionally conforming, therefore a cut failure as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was functionally conforming.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key11688828
MDR Text Key247279864
Report Number1644019-2021-00274
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657524372
UDI-Public380657524372
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number8065752437
Device Lot Number2275739H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2021
Initial Date FDA Received04/19/2021
Supplement Dates Manufacturer Received07/01/2021
03/01/2022
Supplement Dates FDA Received07/02/2021
03/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-