ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)
|
Back to Search Results |
|
Catalog Number 8065752437 |
Device Problem
Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A doctor reported that the probe did not cut and dragged vitreous during a procedure.The product was replaced and procedure completed with no patient harm.
|
|
Manufacturer Narrative
|
A sample was not received at the manufacturing site for evaluation for the report of the cutter was not working and it was dragging the vitreous; therefore, the condition of the product could not be verified.The product was processed and released according to the product¿s acceptance criteria.A sample was not received at the manufacturing site and the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
One opened probe was received, with a tip protector, in a tray along with other items.The sample was visually inspected and found to be nonconforming with orange/brownish foreign material on the port face.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for both functional test.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be functionally conforming, therefore a cut failure as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was functionally conforming.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
|