| Model Number |
1119.5125 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not available for evaluation.Imaging provided confirmed the rod slipping out of the screw head on the left side.The slippage is consistent with increased instability in the construct following the previously reported screw loosening and screw head pop-off.
|
| |
|
Event or Problem Description
|
|
It was reported by a representative from japan that a revision surgery was done for a rod that deviated post-operatively.
|
| |
|
Event or Problem Description
|
|
It was reported by a representative from japan that a revision surgery was done for a rod that deviated post-operatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection of the subject device found no major defects or damage to the rod.The rod had been cut on one end, and there was no curvature placed in the rod.The slippage is consistent with increased instability in the construct following the previously reported screw loosening and screw head pop-off; however, the exact cause of the reported issue could not be determined.
|
| |
|