Model Number 72200568
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that during arthroscopy procedure, inside the patient, the light source 500xl xenon went black and the box was hot.No surgical delay.Unknown how the procedure was finished.Patient injuries were not reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation confirmed the complaint of power up failure.Cause of power failure is a defective power supply/ballast.Unit powered up and passed functional testing with a known good power supply installed.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A complaint history review concluded this was a repeat issue.The complaint of overheating could not be confirmed due to the inability to power up the device in as received condition.Overheating was not observed when the unit was tested with a known good power supply.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.No containment or corrective actions are recommended at this time.The complaint was confirmed.Factors, which could have contributed to the complaint event, include an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11690663
MDR Text Key246315284
Report Number1643264-2021-01464
Device Sequence Number1
Product Code FFS
UDI-Device Identifier03596010643070
UDI-Public03596010643070
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200568
Device Catalogue Number72200568
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/01/2021
Is This a Reprocessed and Reused Single-Use Device? No