H3, h6: it was reported that a patient underwent a first stage revision hip arthroplasty for infection.The polarcup shell (75100442), polarcup insert (75018959) and polarstem (75002115) used in treatment have not been returned for investigation.An evaluation of the complained devices could therefore not be conducted and the reported failure mode could not independently be confirmed.A medical investigation was conducted.It was communicated that no further information would be provided due to lack of consent.Reportedly, infection was the root cause of the reported events; however, no supporting documentation was provided and the ¿type of bacteria¿ was unknown at the time of surgery.The patient impact beyond the reported infection and subsequent 1st stage revision could not be determined.No further medical assessment could be rendered at this time.A review of the complaint history revealed no additional complaint for the batch in question.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported issue.The instructions for use lists infection as a known possible side effect resulting from a hip arthroplasty.A review of the risk management documentation verifies the failure mode and severity of the reported issue.Based on available information the root cause for the reported infection is attributed to a known inherent risk of the devices.However, the executed investigation does not indicate that one of the devices failed to match specification at the time of manufacturing.The need for corrective action is therefore not indicated.Nevertheless, smith + nephew will continue to monitor these devices for similar issues.This complaint will be reopened should additional information or the devices be received.
|