MAUDE Adverse Event Report: VYAIRE MEDICAL AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL AVEA VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number AVEA
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2021
Event Type  malfunction  
Manufacturer Narrative
A vyaire field service representative (fsr) went onsite and evaluated the ventilator.Exhalation sensor connector "o" rings were replaced and secured exhalation flow sensor receptacle.All tests/calibrations okay according to factory specs.Vte is now o.49 ml for 0.50ml dialed vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the avea ventilator experienced low volumes.When set to a certain volume, it appears to be delivering lower.The customer confirmed that there was no patient harm associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEA VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key11694447
MDR Text Key246317910
Report Number2021710-2021-13656
Device Sequence Number1
Product Code CBK
UDI-Device Identifier10846446000641
UDI-Public(01)10846446000641(11)20060720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103211
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAVEA
Device Catalogue NumberAVEA STD COMP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/25/2021
Initial Date FDA Received04/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-