U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. ARTEGRAFT; COLLAGEN VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. ARTEGRAFT; COLLAGEN VASCULAR GRAFT Back to Search Results
Model Number AG636
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An artegraft collagen vascular graft was returned for evaluation.The lot number of the product culture tube returned for evaluation differed from the initial lot number reported (reported lot 20k327-036 and returned lot 20j315-008).Clarification for the correct lot number related to the complaint allegation and the lot number of the product returned was requested from the complainant; to date no response was received.Additional clarification was also requested for the customer statement "the balloon did not inflate" as the artegraft collagen vascular graft product does not contain a balloon.The returned product was visually inspected and pressure tested per release criteria.The evaluation indicated that the graft length did not match the release length of either possible lot number provided.The edge of the graft was not a straight cut and was likely trimmed by the customer after opening.The returned product graft was stuck to lid inside the tube (it is unclear how this occurred).Pressure testing of the returned product verified excessive bubble formation at the graft side walls; no tributaries were identified as showing bubbles.Our review of the lot history records for both batch numbers (20k327 and 20j315) did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.During the manufacturing process, the exact graft was inspected by a manufacturing technician for holes in the graft and proper ligation of the tributaries.No defect was noted during the inspection process.All grafts from these product batch numbers were also tested for pressure test.All of the samples met specifications.Further, we have not received any other complaints of a similar nature from either possible product batch, 20k327 and 20j315.Therefore, we believe that it was an isolated incident.We have also reviewed our complaint history records for last 2 years and searched for a similar issue where a hole/tear was found in an artegraft collagen vascular graft; there is no confirmed trend was identified for this issue.We remain inconclusive about the root cause of the issue, but based on the documentation and complaint history review, we do not believe there is a systemic issue with these devices.No corrective action is needed at this time.The issue was identified during pre-implant preparation and the graft was not utilized; there was no reported patient adverse event.The complaint issue will continue to be monitored within quality assurance trending and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
A lemaitre sales representative received an email notification from the hospital stating "recently the staff was opening an ag636 artegraft for a case and they said it was leaking and the balloon did not inflate.From my understanding it was not implanted in the patient." no patient adverse events were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTEGRAFT
Common Device Name
COLLAGEN VASCULAR GRAFT
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
206 north center drive
north brunswick NJ 08902 4247
MDR Report Key11697296
Report Number2247686-2021-00001
Device Sequence Number14224159
Product Code LXA
UDI-Device Identifier00316837000237
UDI-Public(01)00316837000237(17)231031(10)20J315-008
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
N16837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberAG636
Device Catalogue NumberAg 636
Device Lot Number20J315-008
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2021
Initial Report FDA Received Date04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age33 YR
-
-