| Model Number |
2800 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Gradient Increase (1270)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Stenosis and/or regurgitation, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis and/or regurgitation.Multiple requests for additional details regarding this event (i.E., medical records) have been performed; however, no additional information has been provided.The subject device is also unavailable for evaluation as the it remains implanted in the patient.Therefore, the reported event could not be confirmed.There is currently insufficient information to determine the root cause of this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If any additional information is received, a supplemental report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 17 years, seven (7) months due to stenosis and regurgitation.The tavr was performed with a 20mm 9750tfx transcatheter valve.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 17 years, seven (7) months due to stenosis and regurgitation.The tavr was performed with a 20mm 9750tfx transcatheter valve.There was no allegation that the surgical valve failed or malfunctioned prematurely.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: the device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.Updated b5.Updated d4 expiration date.Updated h4 device manufacturer date.Updated h6 type of investigation by adding code-3331.H11: corrected data: corrected h6 investigation findings by replacing code-180 with code-3221 based on the additional information obtained, this event is no longer considered reportable, and this correction is being submitted.
|
| |
|