U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 2800
Medical Device Problem Codes Fluid/Blood Leak (1250); Gradient Increase (1270)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Stenosis and/or regurgitation, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis and/or regurgitation.Multiple requests for additional details regarding this event (i.E., medical records) have been performed; however, no additional information has been provided.The subject device is also unavailable for evaluation as the it remains implanted in the patient.Therefore, the reported event could not be confirmed.There is currently insufficient information to determine the root cause of this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.If any additional information is received, a supplemental report will be submitted accordingly.
 
Event or Problem Description
It was reported that a patient with a 21mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 17 years, seven (7) months due to stenosis and regurgitation.The tavr was performed with a 20mm 9750tfx transcatheter valve.
 
Event or Problem Description
It was reported that a patient with a 21mm 2800 pericardial valve, implanted in the aortic position, underwent a valve-in-valve procedure after an implant duration of 17 years, seven (7) months due to stenosis and regurgitation.The tavr was performed with a 20mm 9750tfx transcatheter valve.There was no allegation that the surgical valve failed or malfunctioned prematurely.
 
Additional Manufacturer Narrative
H10: additional manufacturer narrative: the device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.Updated b5.Updated d4 expiration date.Updated h4 device manufacturer date.Updated h6 type of investigation by adding code-3331.H11: corrected data: corrected h6 investigation findings by replacing code-180 with code-3221 based on the additional information obtained, this event is no longer considered reportable, and this correction is being submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key11697797
Report Number2015691-2021-02583
Device Sequence Number3511429
Product Code DYE
UDI-Device Identifier00690103021653
UDI-Public00690103021653
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2003
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2006
Device Model Number2800
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/26/2021
Supplement Date Received by Manufacturer04/26/2021
Initial Report FDA Received Date04/20/2021
Supplement Report FDA Received Date05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
-
-