| Model Number |
APB-3-4-3D-ES |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the axium prime coil failed to detach.The patient was undergoing treatment for a ruptured, saccular aneurysm located in the anterior communicating artery.The max diameter was 4mm, and the neck diameter was 2.5mm.The patient's blood flow and vessel tortuosity were normal.It was reported that when the doctor attempted to detach the coil it could not be cut even when the detacher was used twice.Afterwards, an emergency detachment was attempted, but it could not be detached.Two attempts were made with the instant detacher, and one attempt was made manually.There were no problems with the instant detacher. the coil was replaced, and the patient did not experience any injury.The devices were prepared according to the instructions for use (ifu).
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received from a healthcare provider (hcp), via a manufacturing representative (rep), indicated there were no issues encountered prior to non-detachment, and there was no visible damage observed to the pushwire after removal.The lot number of the instant detacher used was unknown.
|
| |
|
Search Alerts/Recalls
|