| Catalog Number |
ME2010 |
| Medical Device Problem Codes |
Leak/Splash (1354); Difficult to Remove (1528); Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that 60 in non-dehp minibore extension set was difficult to disconnect and the tubing ruptured.The following information was provided by the initial reporter: material #: me2010 batch/lot #: unknown it was reported that they are experiencing issues with the tubing when trying to disconnect.Verbatim: we are experiencing issues with disconnection because the tubing functions differently from our normal tubing.We have had multiple issues with the tubing cracking/breaking when trying to disconnect.I believe the breaks are due to force and lack of knowledge about appropriate disconnection.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: no product or photo was returned by the customer.The customer complaint of tubing cracking/breaking when trying to disconnect could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on model me2010 because a lot number was not provided by the customer.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that 60 in non-dehp minibore extension set was difficult to disconnect and the tubing ruptured.The following information was provided by the initial reporter: material #: me2010 batch/lot #: unknown.It was reported that they are experiencing issues with the tubing when trying to disconnect.Verbatim: we are experiencing issues with disconnection because the tubing functions differently from our normal tubing.We have had multiple issues with the tubing cracking/breaking when trying to disconnect.I believe the breaks are due to force and lack of knowledge about appropriate disconnection.
|
| |
|
Search Alerts/Recalls
|