• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 4.5MM FULL RADIUS PLATINUM SERIES; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 4.5MM FULL RADIUS PLATINUM SERIES; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72203521
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that during the surgery when the shaver was being use, metal shavings were found in the knee joint when oscillate , forward and reverse mode was being used.No delay or other complications were reported.Surgery was completed using the same device.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM FULL RADIUS PLATINUM SERIES
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11698612
Report Number1219602-2021-00925
Device Sequence Number11633100
Product Code HAB
UDI-Device Identifier00885554024333
UDI-Public00885554024333
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203521
Device Catalogue Number72203521
Initial Date Received by Manufacturer 03/30/2021
Supplement Date Received by Manufacturer05/19/2021
Initial Report FDA Received Date04/21/2021
Supplement Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-