RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ELEGANCE WITH SSR 230V50HZ, EU; COMPRESSOR, AIR, PORTABLE,
|
Back to Search Results |
|
| Model Number |
1099973 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received this device for evaluation from a distributor in (b)(4) for a report of "does not turn on".Upon evaluation, it was noted there was physical damage to the ac power cord, with evidence of exposure to the conducting wires.The exact date the physical damage occurred was not reported by the distributor.The reported fault of "does not turn on" was confirmed.The root cause of this failure is typically due to the cord being exposed to forces beyond intended design.The innospire elegance device is intended to provide a source of compressed air for medical purposes.It is to be used with a pneumatic (jet) nebulizer to produce aerosol particles of medication for respiratory therapy for children and adults.Labeling warns the user "it is recommended to have a backup device for respiratory delivery in case a situation arises when your nebulizer can not be used".This device meets all relevant ul and iec standards.This device is not labeled for life support.Based on the information available, the manufacturer concludes no further action is required.
|
| |
|