| Model Number |
7K77-25 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Was this device service by a third party? no.An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The male user of the architect i1000sr reported a splash to the eye with the architect progesterone control.His eyes were flushed with water and saline.The user was wearing a lab coat and gloves, but not eye goggles.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow up is submitted to populate fields d8 and/or h6 with data that had previously been provided in field h10.There is no change to the content of the data.
|
| |
|
Additional Manufacturer Narrative
|
|
Use error may have contributed to the complaint issue regarding the use of personal protective equipment (ppe) while using the product.No protective eyewear as per product labeling was in use at the time of the incident.No trends or similar issues for splashing/exposure as described in this ticket were identified.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.Labeling was reviewed and found to adequately address the biological and chemical safety hazards that may exist including the wearing of personal protective equipment (ppe).Based on the investigation no systemic issue or product deficiency was identified.
|
| |
|