| Model Number |
TN,DXH 500 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The fields service engineer (fse) could not identify a specific malfunction or faulty part associated to this event.There was no repair or replacement of a component performed.The customer did not call back with a recurrence of the event as of (b)(6)2021.Section a patient information was not provided by the customer.Bec internal identifier case: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported that their dxh 500 hematology instrument generated erroneously high platelet (plt) results on one patient with no flagging or alert on plt clumps the patient sample was initially run on a dxh 800 hematology instrument which generated r flags and a plt result of 14.0 and prompted the operator to review the results.They reran the patient on the dxh 500 generating much higher plt results (125.4) prompting for a manual slide review that coincided with the initial run on the dxh 800 and confirmed plt clumps on the sample.There was no indication of erroneous results being reported outside the laboratory raw data from the instrument was requested but not provided.Patient results were provided and assessed for this event.There was no report of change or effect to patient treatment.There was no adverse event reported as a results of this occurrence.
|
| |
|
Search Alerts/Recalls
|