Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2021
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the surgical table.During the technicians inspection, he found evidence of fluid intrusion within the power supply.While onsite, the technician spoke with facility personnel who stated that the table had recently been used in a fluid intensive procedure.The fluid intrusion caused the power supply to short resulting in the reported event.The 4085 surgical table is designed to meet ipx4 fluid ingress standards, and the normal fit of the covers, along with the rtv sealant that is applied, will prevent fluid intrusion.In the event the amount of fluid present during the procedure exceeds the ipx4 rating, it is the user facility's responsibility to ensure the table is properly draped and/or a fluid catchment device is utilized to limit the amount of fluid that may come in contact with the 4085 surgical table.The table was installed in 2018 and is not under steris service agreement; the user facility's third-party service provider is responsible for all maintenance activities.The 4085 surgical table operator manual states (5-3), "some procedures performed on table mattresses or pads may involve excessive fluid exposure.Be sure to drape accordingly." the technician made the necessary repairs, tested the table, confirmed it to be operating according to specifications, and returned it to service.A steris account manager offered in-service training on how to properly drape the 4085 surgical table; however, the user facility declined this offer.No additional issues have been reported.
 
Event Description
The user facility reported that their 4085 surgical table stopped working at the beginning of a patient procedure.The patient was transferred to a different table and the procedure was completed successfully.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4085 SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key11701172
MDR Text Key260247963
Report Number1043572-2021-00027
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/26/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-