MAUDE Adverse Event Report: ANGELINI THERMACARE MUSCLE/JOINT 8HR 3+1CT (BONUS); HOT OR COLD DISPOSABLE PACK.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGELINI THERMACARE MUSCLE/JOINT 8HR 3+1CT (BONUS); HOT OR COLD DISPOSABLE PACK. Back to Search Results
Model Number ThermaCare HeatWraps - Muscle & Joint
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Codes Pain (1994); Skin Tears (2516); Partial thickness (Second Degree) Burn (2694)
Date of Event 03/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The root cause and root cause sub class cannot be identified.There was limited device-specific information provided, not batch number or return sample was available for evaluation.Without a batch reference number or a return sample, a manufacturing and technical evaluation cannot be completed for the wrap involved in this case.There is not a product quality-related trend identified for the subclass adverse event safety request investigation.The manufacturing operation employ quality control procedures, which include in- process testing, thermal testing, and visual inspection, to ensure the quality of the packaged product.This is an adverse event for a burn and misuse of the product.A risk calculation cannot be determined as there is no reasonable suggestion of a device malfunction.
 
Event or Problem Description
On (b)(6) 2021, a spontaneous report from the united states was received from a consumer regarding a (b)(6) female (ethnicity was not specified) who was using the thermacare muscle/joint 8hr 3 +1 ct(lot number: lcj8289no6/02; expiration date may-2020).The consumer declined to provide medical history and information about usage of any concomitant products.The consumer had used the product for many years without issue.On (b)(6) 2021, the consumer applied one thermacare muscle/joint heat wrap topically to her lower back and then she went to sleep.Six hours later, the consumer woke up to a burning sensation.She immediately removed the product.When she removed the product, she ripped off some skin.She thought the pain was due to the fact that the patch had too much adhesive.However, she had her mother (who is a burn nurse) look at the area and she stated there was a burn at the application site.Initially, no treatment was performed except she put a loose fitting shirt.On (b)(6) 2021, the application site area was painful and a blister had broken open.On (b)(6) 2021, the consumer went to an emergency department and was diagnosed with a second degree burn.For treatment she was given a neosporin mesh as a dressing to protect the area until she had her follow up with her primary medical doctor on (b)(6) 2021.On (b)(6) 2021, the consumer visited her primary care physician and was given oral oxycodone and tramadol for pain control.She was given silvadene (silver sulfadiazine) for treatment of the burn area.She has been having dressings applied to her the area by her mom.She told the doctor about the use of thermacare but the doctor did not tell her why he thought the product would have caused a burn.As of (b)(6) 2021, the consumer's symptoms have not improved but, have not became any worse.She was told that she may need to be referred to a wound care for management, but she has not heard from her provider as to whether they could get her an appointment for assessment.Currently, she does not have any follow up scheduled with her primary care provider.She was upset that the product was still on the market.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMACARE MUSCLE/JOINT 8HR 3+1CT (BONUS)
Common Device Name
HOT OR COLD DISPOSABLE PACK.
Manufacturer (Section D)
ANGELINI
1231 wyandotte dr
albany GA 31705
MDR Report Key11703146
Report Number3007593958-2021-00004
Device Sequence Number3832059
Product Code IMD
UDI-Device Identifier00305733013144
UDI-Public00305733013144
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2022
Device Model NumberThermaCare HeatWraps - Muscle & Joint
Device Lot NumberLCJ8289NO6/02
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2021
Initial Report FDA Received Date04/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient Weight131
-
-