• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16B-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16B-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 52-1465
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2021, the revision surgery was performed.In the revision, after the head was removed, the surgeon tried to remove the stem, but the stem couldn¿t be removed, and it was left in the patient.The non-depuy implants were set, and the surgery was completed within 30 minutes delay.The surgeon commented that the cold welding occurred between the stem and sleeve and the working space was narrow, and he gave up using the schneider cage due to this reason.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds LS11 8DT
UK   LS11 8DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581
6107428552
MDR Report Key11703897
Report Number1818910-2021-08419
Device Sequence Number3460799
Product Code LPH
UDI-Device Identifier10603295170624
UDI-Public10603295170624
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number52-1465
Device Catalogue Number521465
Device Lot Number8018457
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 03/31/2021
Supplement Date Received by Manufacturer06/07/2021
05/13/2022
Initial Report FDA Received Date04/21/2021
Supplement Report FDA Received Date06/09/2021
05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SROM 9/10 16X11X130 30
Patient Age71 YR
Patient SexFemale
Patient Weight68 KG
-
-