| Model Number |
657324S |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of the investigation any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported the screw lost a metal part/splintered while fixing the plate on the bone.
|
| |
|
Event or Problem Description
|
|
It was reported the screw lost a metal part/splintered while fixing the plate on the bone.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed based on the damage pattern shown at the locking thread.The received screw was found to be damaged with its locking threads sheared off.The threads look absolutely splayed which could explain the loss of the angular locking ability.The hex at the head of the screw shows severe signs of deformation.These damages indicate towards excessive force that was applied to screw in the screw with some mis-alignment/angulation as well.A potential non-conformity report was initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No deviation from the specifications was noted.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.
|
| |
|
Search Alerts/Recalls
|