• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L24MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L24MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 657324S
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported the screw lost a metal part/splintered while fixing the plate on the bone.
 
Event or Problem Description
It was reported the screw lost a metal part/splintered while fixing the plate on the bone.
 
Additional Manufacturer Narrative
The reported event could be confirmed based on the damage pattern shown at the locking thread.The received screw was found to be damaged with its locking threads sheared off.The threads look absolutely splayed which could explain the loss of the angular locking ability.The hex at the head of the screw shows severe signs of deformation.These damages indicate towards excessive force that was applied to screw in the screw with some mis-alignment/angulation as well.A potential non-conformity report was initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No deviation from the specifications was noted.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L24MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11704193
Report Number0008031020-2021-00177
Device Sequence Number14322942
Product Code HWC
UDI-Device Identifier07613327086997
UDI-Public07613327086997
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number657324S
Device Catalogue Number657324S
Device Lot NumberD44320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Initial Date Received by Manufacturer 03/26/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/21/2021
Supplement Report FDA Received Date05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-